Bio-medical
    
Bio-Medical
Stat Fax® 4200
• Fully automatic reading.
• Reads 96 wells in about 12 seconds.
• Interactive touch-screen LCD with USB mouse option.
• Bichromatic optics. Standard model includes four filters, six filters are available for a nominal charge.
• On board data reduction calculations include regressions, curve fitting and cutoff absorbance calculations with step-by-step user prompting.
• NRTL and CE mark.
• IAD filters
 

 

Stat Fax®4200

Linear Measurement Range

0.0 to 4.0 Absorbance units (A)

Photometric Accuracy

+/- 1%

Stability

Drift of no more than 0.005A in 8 hours

Light Source

Halogen lamp with lamp saver feature

Wavelengths

Standard: 405, 450, 492, and 630nm
6-fiter: 405, 450, 492, 545, 600, and 630nm
Filter options from 405nm to 700nm

Filter Type

IAD hardcoat interference, 10nm half bandpass

Vessel

Standard 96-well microtiter plates, or strip trays

Display

Interactive 5.7” (14.5cm) graphical LCD with touch-screen

Printer

Thermal graphics

Interface

USB mouse

Power Requirements

115V or 230V AC, 50-60Hz (auto-sensing)

Speed

Reads absorbances of 96 wells in about 10 seconds

Calculation Modes

Single point calibration, point-to-point curve fit, cutoff absorbance, polynomial regression, linear regression (log & linear), multipoint % absorbance

Additional Features

User-programmable open system, selectable plate formatting, alphanumeric test naming, automatic interpretation options, duplicate well options, curve plotting and editing, flags and error messages,

Plate Transport

Stepper motor

Enclosure

Flame-retardant ABS plastic cover and base

Dimensions

Approx. 7.5 x l3 x 16.5 in (19 x 20.3 x 42cm) weighs 15 lbs. (6.8 kg)

 