Bio-medical
    
Bio-Medical
Stat Fax® 4500
• Accommodates 12mm tubes or cuvettes.
• Interactive touch-screen LCD with USB mouse option.
• Bichromatic optics. Standard model includes six filters - 340, 405, 505, 545, 580, and 630nm. Alternate filters available from 340nm - 700nm.
• On board curve fitting calculations with step-by-step user prompting.
• Single point calibration by standard or factor, multipoint calibration, kinetic rate by standard or factor (batch or singly).
• NRTL and CE mark.
• Stat Fax® readers use IAD filters.
 

 

Stat Fax®4500

Linear Measurement Range

-0.0 to 3.0 Absorbance Units (A)

Photometric Accuracy

+/- (1% of the reading + 0.005 A)

Stability

Drift of no more than 0.005A in 8 hours/bichromatic

Light Source

Tungsten lamp with lamp saver feature

Standard Wavelengths

340、405、505、545、580、630

Type of Filter

IAD hardcoat interference, 10nm half bandpass

Tube Size

12 mm round is standard

Display

Interactive touch-screen 3.5” LCD, color graphic display

Printer

Thermal dot matrix, with graphic capability

Power Requirements

115V or 230V AC, 50-60 Hz (universal input)

Speed

Reads, calculates and prints results in 3 seconds/tube

Calculation Modes

Single point calibration by standard or factor, multipoint calibration with point-to-point curve fit, rate by standard or factor (batch or singly)

Test Menu

More than 50 open channels to store tests. Stores all parameters including wavelengths, calculations, unit codes, linear and normal ranges, rate timing, standard values, test names, and previous standard curve.

Temperature

Read cell and incubation block, user selectable

Enclosure

Painted flame-retardant ABS plastic cover with metal base.

Dimensions

Approximately 9 x 13.5 x 5 in. (24 x 34 x 13 cm) 10 lbs. (4.5 kg)

Certifications

NRTL listed, CE Mark

 
index